The product has been added to the request form. You can check its content by clicking on the ‘view your request’ below the left product menu. You can continue adding other items.
BTL-08 LT ECG
12-channel electrocardiograph with 5.7'' colour touch screen and A5 print
- Colour touch screen for comfortable operation
- 210mm print on ECG paper (roll, fax, folded)
- Direct connection and print on laser printer - via USB port, independent of PC
- Pacemaker detection
- Upgradeable to a PC data management and stress-test system
| Ordering no: | |
|---|---|
| C08LT.001v300 | |
| C08LT.002v300 | with diagnostics |
technical parameters
| Dimensions | 407 x 312 x 125mm |
| Weight approx. (less accessories) | 5.9kg (LC) / 5.7kg (LT) |
| Dimensions of display | 171 x 128mm (LC) / 120 x 89mm (LT) |
| Screen resolution | 640 x 480 dots (LC) / 320 x 240 dots (LT) |
| Mains supply | 100V-230V, 50-60Hz |
| Frequency response | 0,04Hz-150Hz |
| Digital resolution | 3,9uV |
| A/D conversion | 13bits |
| Sampling frequency | 2kHz/12 leads, 18 kHz/1 lead |
| Dynamic range | 15,9mV |
| Polarisation voltage | 400mV |
| Max. consistent voltage | 5V |
| Input impedance | > 20MOhm |
| Common mode rejection | > 100dB |
| Safety standards | IEC 601-1, IEC 601-2-25, IEC 601-1-2, IEC 601-1-4, ISO 14971 |
| Accumulator capacity | 2-3 hour continuous recording, 35m continuous printing |
| Charging time | approx. 6 hours (if completely discharged) |
| Protection class | II according to IEC 536 |